사양
Tolerance % E36105B
DC output rating (0 to 40°C)
Max. voltage : 60 V
Max. current : 0.6 A
Load regulation ± (% of output + offset)
Voltage : < 0.01%+ 10 mV
Current : < 0.02% + 30 μA
Line regulation ± (% of output + offset)
Voltage : < 0.01%+ 7 mV
Current : < 0.02% + 30 μA
Output ripple and noise (20 Hz to 20 MHz)
Voltage RMS 1.55 mV
Pk-Pk 30 mV
Accuracy 12 months (23°C ± 5°C)
Programming accuracy ± (% of output + offset)
Voltage : 0.05% + 20 mV
Current : 0.05% + 0.4 mA
Readback accuracy ± (% of output + offset)
Voltage : 0.05% + 12 mV
Current : 0.05% + 0.3 mA
Low range current : 0.25% + 40 μA
(0-3 mA)
Load transient recovery time (Time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load)
Voltage settling band : 150 mV
Time : 50 μs
Resolution
Program (Average)
Voltage : 3.6 mV
Current : 36 µA
Readback
Voltage : 2.4 mV
Current : 24 µA
Low range current : 180 nA
Program (Meter) Minimum perceivable change
Voltage : 3 mV
Current : 1 mA
Readback (Meter)
Voltage : 3 mV
Current : 1 mA
Low range current : 1 µA
Output ripple and noise (20 Hz to 20 MHz)
Current : 200 μA
Overvoltage protection (OVP) ± (% of output + offset)
Accuracy : 1 V
Activation time (average time for the output to start to drop after OVP or OCP condition occurs)
Overvoltage (OVP) : < 1.5 ms when the trip voltage is greater than or equal to 3 V
Overcurrent (OCP) : < 1.5 ms
Command processing time : < 10 ms
Programming temperature coefficient per ºC (% of output + offset)
Voltage : 1.8 mV
Current : 60 µA
Readback temperature coefficient per ºC (% of output + offset)
Voltage : 60 µA
Current : 30 µA
Remote sense (max. voltage in load lead)
Output can function as described with up to a 1-V drop per load lead
Up/down programming settling time to within 1% of total excursion
Up, full load : 50 ms
Up, no load : 50 ms
Down, full load : 30 ms
Down, no load : 250 ms
I/O interfaces : LAN (LXI Core) and USB 2.0 FS (TMC488)